Mini Photoelectric Spectrometer for photoelectric properties examination

Photocurrent and OCP measurements under illumination from LED Revolver.

description

Mini Photoelectric Spectrometer setup allows to apply various measurement techniques: chronoamperometry, voltammetry, cyclic voltammetry and the light intensity modulated photocurrent/photovoltage. Those techniques enables to examine the photoelectric dynamics of the sample. Samples may be investigated using two- or three-electrodes configuration.

Quantities measured by the instrument:

  • I-V characteristics of a sample under illumination,
  • Open Circuit Potential (OCP) as a function of light wavelength,
  • Photocurrent as a function of light intensity,
  • IPCE or EQE at wavelengths provided by the LED Revolver,
  • Surface Photovoltage (SPV),
  • Space Charge Region capacitance if a super-bandgap wavelength LED is present in the LED Revolver.

Measurement techniques:

  • Chronoamperometry,
  • Voltammetry,
  • Cyclic voltammetry,
  • Chopped illuminance and more.
Click the photo to enlarge

specifications

LED Revolver

  • 10 LED slots,
  • Default LED set: 365 nm, 385 nm, 405 nm, 450 nm, 520 nm, 590 nm, 630 nm, 740 nm, 840 nm, white (3000 K),
  • Custom LED selection available,
  • Typical electric power of LEDs: 3W,
  • Light intensity modulation: square, sinusoidal, triangle or ramp waveform with adjustable amplitude and bias light intensity,
  • Light modulation frequency range: 1 µHz ÷ 10 kHz,

Potentiostat

  • Bias potential range: -5 V ÷ 5 V,
  • Current ranges: 10 nA, 100 nA, 1 µA, 10 µA, 100 µA, 1 mA, 10 mA,
  • Current resolution: from 1 pA to 100 nA (range dependent),

Controller

  • Maximum sampling rate: 1 kSample/s,
  • Connection with PC: via USB 2.0,

Electrochemical cuvette

  • Adapted for transparent and opaque samples,
  • Sample may be illuminated through the electrolyte or through the rear optical window,
  • PTFE container for electrolyte,
  • Reference electrode: Ag/AgCl,
  • Counter electrode: Pt.

The instrument may be extended with the Sample Chamber to screen the ambient light and electromagnetic fields.

exemplary results

I-V characteristics

kliknij obrazek, aby powiększyć

OCP versus light intensity

kliknij obrazek, aby powiększyć

3D map of I-V curves for different light intensities

kliknij obrazek, aby powiększyć

 

Software
FotoGUI Zobacz
ACCESSORIES The following accessories are compatible with the product
Dummy cell
LED revolver
Light Calibrator
Platinum electrode
Potentiostat
Silver chloride reference electrode
Sample Chamber
Universal cuvette
CONTACT US
LEAVE YOUR CONTACT INFORMATION - We will respond within 48h





    I agree that provided personal information will be processed by Instytut Fotonowy Sp. z o.o. only for the purpose of business offer for the relevant product.

    CONTACT US - We will address all your questions + 48 733 003 380 office@fotonowy.pl

    Similar products