Single-Point Kelvin Probe to investigate the Work Function

High precision, non-contact Work Function determination for semiconducting and conducting materials.

description

The Kelvin Probe system allows to precisely measure the Contact Potential Difference (CPD) and to determine the Work Function of conducting and semiconducting material.
The non-contact technique of CPD measurements employs the Probe Tip-oscillating reference electrode made of a golden mesh. The examined sample serves as the second electrode in the circuit. The Work Function value is evaluated indirectly from the measured CPD of the sample referenced to the gold tip.
The Kelvin Probe system may be used in various types of studies of the material surfaces such as: corrosion and photocorrosion process, adsorption/desorption, surface charging, catalytic activity and others. Subsequent CPD measurements in adjustable time intervals allows to monitor the changes in real time. The measurement method is accurate for many types of samples: organic and inorganic semiconductors, solar cells, photoactive materials. During the measurement, the instrument is covered with the Faraday Cage to screen the ambient light and electromagnetic fields.

Kelvin Probe system enables to investigate:

  • The Work Function of a material,
  • Position of Fermi level with respect to the conduction and valence band edges,
  • The type, n or p, of a semiconductor,
  • Energy gap of a semiconductor,
  • Surface potential (SPV by flat band technique),
  • Surface adsorption and chemisorption effects on the Work Function,
  • Information on the surface states density,
  • Surface charging and discharging effects,
  • Minority carriers diffusion length,
  • Carrier recombination rates (especially for indirect bandgap semiconductors),
  • Information on buried interfaces,
  • Temperature influence on the Work Function,
  • Humidity influence on the Work Function.

kelvin probe modules

The Kelvin Probe set consists of:

  • Controller unit,
  • Kelvin Probe instrument with the Probe Tip,
  • Faraday Cage.
click the picture to enlarge

 

The Kelvin Probe instrument is equipped with:

  • Protective bay to park the Probe Tip after experiment,
  • Laser pointer to indicate the area directly,
  • Light source that helps to handle the sample when the instrument is covered with the Faraday Cage,
  • Laser barrier to precisely detect the Probe Tip distance to the sample,
  • Golden Probe Tip,
  • Sample holder.

click the picture to enlarge

Probe Tip

The most important part of the Kelvin Probe instrument is the Probe Tip.  The reference electrode is made of Au mesh of 2.5 mm diameter. It provides high signal to noise ratio, even from a distance of 0.5 mm above the sample. Thus, it does not matter if the examined sample surface is rough or polished. The tip oscillations are generated with an electromagnet.

The Probe Tip is the component fully designed and manufactured by Instytut Fotonowy Sp. z o.o.

click the picture to enlarge

 

Sample holders

The standard Kelvin Probe Set includes two types of sample stands:

  • Bottom contact stand for samples on conducting substrates,

  • Top contact stand for non-conducting substrates,

Faraday Cage

The entire instrument is covered with the Faraday Cage, shielding the setup from the ambient light and electromagentic fields.

click the picture to enlarge

Gas-tight Faraday Cage with inert gas flow system

Scince measurements of the CPD is often sensitive to environmental factors like temperature, humidity, dust and chemical contaminants, Kelvin Probe may be placed inside the hermetic version of the Faraday Cage with the inert gas flow system.

click the picture to enlarge

Laser barrier

The Kelvin Probe is equipped with the Laser barrier system. System automatically detects the sample sub-
strate. Each time the probe approaches the sample being examined, an accurate distance between the sample surface and the Probe Tip is measured with 20 µm precision.

click the picture to enlarge

 

Sample illumination

The sample may be illuminated during the measurements from the optical fiber situated above the Probe Tip. The Probe Tip lets the light through. The light at optical fiber input may be provided by the LED Revolver, Xenon Lamp combined with the Monochromator or any other type of the light source desired.

click the picture to enlarge

 

Several types of liquid lightguides from Lumatec may be used to provide the sample illumination.

Lumatec wave guides types

specifications

General:

  • Weight: 15 kg,
  • Size: 40x40x45 cm,
  • PC connectivity: USB 2.0,
  • Power supply: 230 V, 50 Hz or 115 V, 60 Hz,
  • Measurement technique: 2-channel lock-in-amplifier,
  • Laser barrier to automatically detect the substrate and prevent the tip from running into the sample,
  • Auxiliary sensors: humidity, temperature,
  • Small lamp,
  • Laser pointer for examined point of the surface indication,
  • Sample holders:
    – Freely shaped solid state with top contact holder,
    – Bottom contact holder,
    – Electrochemical holder,

Faraday Cage:

  • Standard / gas-tight with inert gas flow system,

Measurement unit:

  • Bias voltage range: -5 ÷ 5 V,
  • Voltage measurement resolution: 0.15 mV,
  • Current ranges: 300 nA, 30 nA, 3 nA, 300 pA,

Probe Tip:

  • Probe Tip type: Au mesh, 2.5 mm in diameter,
  • Tip positioning resolution in vertical axis: 20 µm,
  • Automated resonance frequency scanning,
  • Adjustable oscillation amplitude,
  • Automated removal of the parasitic current of the tip,
  • Partial transparency for the light,
  • Typical CPD measurement distance: 0.2 – 1 mm,

 

exemplary results

CPD values recorded for a Photodiode. The plot represents the response for a saw tooth type light intensity modulation.

 

click the picture to enlarge

AKCESORIA Przy zakupie produktu warto go wyposażyć w poniższe akcesoria
LED Changer
UV-VIS Radiometer & Photometer
SKONTAKTUJ SIĘ Z NAMI
ZOSTAW KONTAKT DO SIEBIE - Odezwiemy się w przeciągu 48h





    I agree that provided personal information will be processed by Instytut Fotonowy Sp. z o.o. only for the purpose of business offer for the relevant product.

    SKONTAKTUJ SIĘ Z NAMI - Odpowiemy na wszystkie Twoje pytania. + 48 733 003 380 office@fotonowy.pl

    Podobne produkty

    单点开尔文探针

    开尔文探针系统可以高精度、高准确度地精确测量不同半导体和导电材料的功函数

    menu nawigacyjne
    0 z 0
    描述

    许多材料的特性取决于其表面。其中最重要的表面特性之一就是功函数。

    开尔文探针系统可以高精度、高准确度地精确测量不同半导体和导电材料的功函数。

    非接触式测量涉及振荡电容器,其中一个电极是试样,另一个是由金网格制成的振荡参比电极。因此,测量值以金的功函数(5.1 eV)为基准。在足够长的积分时间内,测量分辨率可达 0.1 meV。开尔文探针系统可用于各种表面研究,包括腐蚀、吸附/解吸、表面充电、催化活性等。

    为方便起见,探头配备了激光指示器和数码显微镜。电动垂直定位系统可使参比电极与试样表面保持相同距离(精度为 10 微米)。可选配 X-Y 定位系统。该系统还配备了宽带光导,可与数字控制单色仪和高压氙弧灯连接。波长范围从 300 纳米到 1000 纳米,可用于连续或脉冲样品照明。这种配置可用于测量表面的功函数和电荷、表面光电电压光谱和其他光物理技术,光电电压分辨率优于 5 mV(积分时间长时为 0.1 mV)。整个系统安装在接地的法拉第笼中,以减少电噪声和杂散光的影响。这种表面光电电压光谱仪是研究光催化剂、有机和无机半导体、太阳能电池及其他光活性材料和器件的重要工具。借助这种开尔文探针/表面光电压光谱仪,还可以实时监测半导体的光腐蚀、光吸附和光催化反应。

     

    开尔文探头模块

    该仪器由安装在坚固支架上的振荡金探针、Z 轴电动平台、电子控制单元和法拉第笼组成。

     

    • 样品放置在手动或电动 XY 平台上,两个方向的移动范围均为 1.5 厘米。
    • 探头连接到 Z 平台上,Z 平台可将探头放置在与样品的理想距离上。
    • 红色激光指示器可显示探头在测量过程中摆动的位置。
    • 一旦系统被法拉第笼覆盖,USB 显微镜和白色发光二极管就会发挥作用。
    • 光通过液体波导或光纤传送到样品,因此消除了光源可能产生的电磁干扰。
    • 在空载运行或存放期间,易碎的金探针会被固定在一个受保护的空间内,以防止意外损坏。
    • 系统还配备了自动样品高度传感器。

     

    点击图片放大

     

     

    探头尖端

    探头尖端是完全由 Instytut Fotonowy 设计和制造的组件,它提供:

     

    • 即使距离样品上方 0.5 毫米,也能发出较大信号。 因此,样品表面是粗糙的还是抛光的并不重要。
    • 让光线透过的透明度。 这样样品就可以被垂直光束照亮。

    尖端振荡由电磁铁产生。

    点击图片放大

     

    法拉第牢笼

    整个探头被法拉第笼覆盖,使其免受环境影响。

    点击图片放大

     

    规格

    由于仪器的不断发展,以下参数可能会发生变化

    Voltage range-5 V .. 5 V
    Voltage measurement nominal resolution0.15 mV
    Measurement technology2-channel lock-in amplifier
    Sample holdertranslation stage X-Y, manual (default) or motorized
    X-Y stage stroke15 mm x 15 mm
    Tip typeAu mesh, dia. 2.5 mm
    Tip Z positioningmotorized Z stage
    Positioning with respect to a samplelight barrier
    Tip Z positioning resolution10 µm
    Spectral range for wave guides200 nm .. 2000 nm (depending on a waveguide)
    Liquid waveguides selection (Lumatec)Wybór światłowodów Lumatec
    Auxiliary sensorshumidity and temperature
    Internal Faraday cage illuminationwhite LED
    X-Y positioning aidlaser sample placement indicator
    PC connectivityUSB 2.0
    Size40 x 40 x 45 cm
    Weight15 kg

     

    示范性结果

     

    以下结果来自 Radecka 教授小组(波兰克拉科夫 AGH)。

    抗蚀剂样品的表面光电压谱。光电压是根据脉冲照明与金栅参比电极的表面电位变化来测量的。

    由于测量是在环境空气中进行的,因此仪器会记录空气湿度和温度。这两个参数可能会影响样品的功函数。

    点击图片放大

    AKCESORIA Przy zakupie produktu warto go wyposażyć w poniższe akcesoria
    LED Changer
    UV-VIS Radiometer & Photometer
    SKONTAKTUJ SIĘ Z NAMI
    ZOSTAW KONTAKT DO SIEBIE - Odezwiemy się w przeciągu 48h





      I agree that provided personal information will be processed by Instytut Fotonowy Sp. z o.o. only for the purpose of business offer for the relevant product.

      SKONTAKTUJ SIĘ Z NAMI - Odpowiemy na wszystkie Twoje pytania. + 48 733 003 380 office@fotonowy.pl

      Podobne produkty

      71703.40364.1696